Failure of complex salt spray corrosion test chamber and its treatment method The salt spray test and salt spray aging test of the compound salt spray corrosion test box due to the long-term material due to harsh environment, resulting in some failures, in order to allow users to quickly solve The problem can continue to make the salt spray tester operate normally. We introduce some common faults and treatment methods.
First, the cause of temperature failure: 1. Because the laboratory temperature controller temperature is set too low, you can set the temperature controller at the desired temperature; 2. Because the laboratory safety switch is set too low, you can protect the security Switch setting temperature increases; 3. Heating system failure, electromagnetic (electrical) failure, controller failure, can consult our after-sales department for debugging and repair.
Second, saturated barrel temperature failure: 1. When the temperature of the saturated tank temperature controller is too low, the temperature controller can be set to the desired temperature; When the saturated bucket safety switch is set too low, the compound salt spray corrosion test box can increase the setting temperature of the safety switch. 3. Heating system failure, electromagnetic (electrical) failure, controller failure, can consult our after-sales department for debugging and repair.
Third, the lack of spray failure causes: 1. When the spray regulator is placed too low, you can adjust the spray regulator; 2. Preheat the glass filter block the plug, you can clean the glass filter; 3. When the pressure is set too low, the pressure regulator can be adjusted to a pressure of 1KG per square centimeter. The air compressor is marked with a regulating valve and adjusted to a pressure of 2 KG per square centimeter.
Fourth, when the water shortage warning light is on, it means that the water level is too low. It is necessary to check whether the water level in the laboratory or saturated tank is sufficient.
5. When the air compressor is not in operation, the air compressor itself has the function of self-enveloping. At this time, it can be used as usual, without affecting the test process.
6. When the water tank is not operating, when the water level in the warming water tank is too low, the operating power will be cut off, and the water level in the heated water tank can be added to normal conditions.
VII. Composite Salt Spray Corrosion Test Chamber When the temperature indicator is faulty or the temperature sensor is faulty, you can consult our after-sales department for debugging and repair.
Quilting Mattress Protector,Machine Quilting Design,Equipment for Makingprotection Filter
Sichuan Kisae sewing machine Co.,Ltd. , https://www.kissew.com